Products
The PL-SLP is a very lightweight and small recorder with 16 analog channels and a serial interface to connect a NONIN finger clip sensor.

- Channels:
– 3x EEG (referential)
– 2x EOG (referential)
– 1x ECG (bipolar)
– 3x EMG (bipolar)
– 3x Body position/movement (internal acceleration sensors)
– 2x Breathing
– 1x Flow (mouth and nose)
– 1x Microphone (snoring)
- Digital marker channel for event registration
- Optional NONIN finger clip sensor for:
– 1x Plethysmogram (finger blood pressure)
– 1x Oxygen saturation (percent)
– 1x Pulse rate (calculated from finger pulse signal)
- Further options:
– Sensor for registration of environment temperature and humidity
– Higher Processor frequency for higher sampling rates – up to 1600 Hz per channel
– Display showing operation modes and measuring data
– Wireless recorder operation interface (Bluetooth) with PDA, PC or notebook
- Data are recorded to a Compact Flash Card
- Sleep stading and further quantitative analyses can be applied after data transfer. In combination with patient event markers, data quantity effectively can be reduced to a minimum.
- Using PL-Winsor, irrelevant data segments easily can be cut, and data can be archived and managed together with standard EEG records.
Dimensions
- Size: 12 x 6.5 x 2.2 cm
- Weight: 170 g (including rechargeables and memory card)
Channels
- 3x EEG (referential)
- 2x EOG (referential)
- 1x ECG (bipolar)
- 3x EMG (bipolar)
- 3x Body position/movement (internal acceleration sensors)
- 2x Breathing
- 1x Flow (mouth and nose)
- 1x Microphone (snoring)
- Event marker channel
Technical data of EEG channels
- Amplification rate 615
- Input range ± 2 mV
- Input noise 1.2 µVpp
- Input impedance > 100 MΩ
- Common mode rejection ratio > 77 dB (@50 Hz)
- Frequency range 0.5 ... 60 Hz
- Time constant 0.33 s
A/D conversion
- Two eight-channel A/D-converters
- Resolution 16 Bit (65536 steps) = 61 nV / Bit
- Sample rate can be set by software (128 or 256 Hz; optional 1600 Hz)
Data storage
- Compact Flash Cards (512 MBytes to 4 GBytes)
- By default, about 15 MBytes data per hour are recorded, i.e. up to 34 hours can be stored on a 512 MB Flash Card.
Power Supply
- Two AA rechargeables (NiMH) or batteries
- Power consumption from 0.5 mA (sleep mode) to 30 mA max.
Operating time
- Up to 60 hours with one set of rechargeables
Connectors
- Three connectors for electrode cables
- One connector for serial port (connection to PC, PDA or notebook via optical isolated interface)
Accessories
- Two sets of rechargeables, charger and tester
- Compact Flash Card 512 MB
- Optical isolated serial interface (incl. USB adapter)
- Card reader (with USB interface)
- Carrying pouch for the recorder
- Transport case